5.0kW/100mA Mobile Veterinary X-ray Machine YSX100VET
1. The power supply requirements 1.1 Voltage requirements: single phase line voltage to 220 + 10% for VAC, 50 Hz 1.2 Input electric power: 7.5 KVA standard ground wire or less
2. Generator 2.1 Inverting frequency: 60 KHZ or higher; 2.2 Filament frequency: ≥20 KHZ; 2.3 Focus: 1.5 mm 2.4 Max. output power: 5 kw 2.5 Radiography tube voltage range: 40 ~ 120 kv; 2.6 mAs range: 1 ~ 200 mas 2.7 Maximum tube current: 100 ma
3 Mechanical requirement 3.1 X ray tube focus range of vertical ground movement: 500 ~ 1750mm or higher; 3.2 X ray source component vertical rotation around the cross arm: + 90°or higher; 3.3 The limit of beam splitter spinning the ball tube window center around: p + 90°; 3.4 Radiography size: 350 mmx350mm or greater (at the SID = 65 cm);
Main functions
4.1 APR animal anatomy program function; 4.2 8 inches true color touch-screen operation interface; 4.3 Malfunctions auto-detecting;
4.4 Automatic calibration parameters;